DAC Premium and DAC Premium Plus use the same double-walled sterilization chambers (twin chamber technology) as hospital autoclaves. The chambers are preheated with steam, which is then used for sterilization. You save time, energy costs and both you and your patients are protected by the best possible sterilization result in a reliable and legally compliant manner. The twin chamber technology also offers a decisive advantage because the air is very quickly evacuated from the hollows of the instruments, sterilization packages and chamber. With autoclaves of the DAC Premium class, you can quickly and easily step into the traceability of the instruments used on patients.
Twin-chamber technology:
Simple installation as a standalone device
Complies with all relevant norms for legal certainty
Intuitive and simple operation via the color touch display
Documentation with printer, CF card or directly in the PC without addition softwareis possible
Short sterilization cycles (e.g. quick program B in 20 minutes including drying)
All-round simplicity:
With the extra-large color touch display, the appearance of the DAC Premium autoclaves also makes them stand out from conventional class B sterilizers. The menu navigation is so intuitive that operator errors are a thing of the past. You can also input any desired settings quickly and easily, or, for example, customize the display background to your tastes.
Installation prerequisites :
Electrical power supply : 90-120V AC?/?190-240V AC, 50?/?60 Hz
Power consumption : 1.1 kW
Compressed air connection :
Input pressure: 5.0 - 8.0 bar; max.
Short-term air consumption: approx. 60 NI/min. at 5 bar
Oil tank capacity : 0.2 l
Water tank capacity : 1.8 l
Water consumption : Approx. 600 ml per cycle (Standard program)
Water quality : < 3 µS
Height open/closed (with Standard lid) :53 cm?/?35 cm x 38 cm x 37 cm
Equipment: unwrapped/solid : 6?/?5 instruments
Weight : 23 kg
Minimum distance from wall/ceiling :10 cm?/?70 cm
Documentation :
RS 232 connections : Printer, documentation software, USB adapter,memory sticks, memory cards, etc.
Printout : Selected sterilization program, date, time, batch number, vacuum, pressure, temperature, serial number, deviation, etc.
Programs :
Sterilization program : Standard program
Selection options : 134°C, 3 min. sterilization
Cycle times may vary depending on the working environment
in which the autoclave is operated. :
overall cycle time: approx. 18 min.
Standard prion program 134°C, 18 min. sterilization overall cycle time: approx. 32 min.
Program for solid instruments
Individual selection of standard, sensitive or prion program,but does not include lubrication.
The cycle times remain the same.
Disinfection program : Program for ultrasonic scalers, ultrasonic tips and syringe nozzles 134°C, 3 min. disinfectionoverall cycle time: approx. 22 min.
Techinal data:
Reference numbers:
65 26 920
65 36 929 (AUS)
Electrical connection : 3,400 W / 220 – 240 V / 50/60 Hz
External dimensions (W x H x D) : 47 x 56.5 x 57.5 cm
Sterilization chamber (diameter x depth) : 25 x 35 cm
Weight (without load) : 59 kg
Chamber volume : 18.4 l
Max. loading : Instruments: 6 kg or textiles: 2 kg
Water tank (aqua dem/aqua dest) :
Fresh water side: 5 l (approx. 7 cycles)
Waste water side: 3 l
Operating time without drying) Drying time
Load up to 6 kg of unwrapped instruments Load up to 2.5 kg of textiles